深圳新光電子稱天平為您提供維修新光珠寶天平vibra黃金電子秤維修克拉秤。維修新光珠寶天平 vibra 黃金電子秤維修 克拉秤ct603ce ct1602ce 電話
故障現(xiàn)象錯誤代碼::
不顯示 顯示虛線 數(shù)字不穩(wěn)定 亂跳 顯示不完整 顯示不清楚 傳感器損壞斷裂 更換簧片彈片 更換音叉 秤盤支架損壞
電子秤支腳 外殼 面板壞按鍵不靈 不能開機 不能置零/去皮 無法轉換單位 元件燒黑燒焦 cpu error 程序丟失 不能稱重 數(shù)字不動
規(guī)格 - 電子天平
可讀性 1 mg
重復性典型 0.5 mg
小稱量值 (usp),典型值 1 g
秤盤外形尺寸(深x寬) 127 mm x 127 mm
線性誤差典型值± 2 mg
---交易 no
重復性校驗砝碼 6 mg (200 g)
穩(wěn)定時間 1.5 s
尺寸 (高x寬) 363 mm x 199 mm
校正 內部/profact
稱量臺尺寸(深x高x寬) 96 mm x 257 mm x 194 mm
物料號 (s) 30115733
possible remediation
there is no indication
on the display.
the ac adapter is not connected. *** check that the ac
adapter is connected (8).
the display is unstable.
[m] remains flashing
without changing.
the balance is subject to air currents or
vibration.
the balance is situated on an unstable
surface.
an object is contacting the sample being
measured, the measuring pan,/the tare.
*** check precautions on
use (2–4).
weight indication
contains an error.
an error was made in the tare subtraction
procedure.
the adjusters remain lifted, resulting in an
incorrect level.
the indication values are inconsistent
after long hours of use,/because the
balance has been moved to a new
location.
*** review the tare
subtraction (10).
*** check the level (8).
*** execute span adjustment
on the balance (17).
[o-err] appears before
the capacity is reached.
gross weight exceeded the capacity of
the balance (weight range = container +
weight of sample).
a section of the mechanism is damaged.
*** check the total weight.
*** execute tare subtraction
again.
*** contact our local dealer.
[u-err] is displayed.
a foreign object is caught between the
measuring pan (pan base)&the balance.
a section of the mechanism is damaged.
*** remove the measurement
pan&examine the
surface beneath it.
[b-err] is displayed.
[d-err] is displayed.
the balance is exposed to static
electricity/noise.
the electrical system of the balance is
malfunctioning.
*** contact our local dealer.
during span adjustment:
[o-err] is displayed.
[1-err] is displayed.
[2-err] is displayed.
a weight heavier than the capacity was
used.
the reference weight is less than 50% of
the capacity.
calibration produced an error of 1.0% or
more.
*** check that the span
adjustment procedure
was performed correctly
(17).
during battery
installation:
the indication
disappears.
[ ] flashes.
no indication is
produced.
the automatic power-off function was
activated.
the battery capacity is low.
*** switch on the power
again. deactivate the
automatic power-off
function, if necessary (13).
*** recharge the battery (25).
*** operate the balance with
the ac adapter.